Inhaltsverzeichnis

Seite 1 - Production

Document/Revision No. IM-GB-PR23-AC-VACCINES: Rev. 1.01Effective: November 1, 2013K-Patents Pharma Refractometer PR-23-ACAppendix to PR-23 Instruction

Seite 2

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201310Appendix to the Instructions Manual PR-232.3 Pharma Refractomete

Seite 3 - Contents

The Pharma Refractometer PR-23-AC2© Copyright K-Patents 2013. All rights reserved.11Figure 2.2 Pharma Refractometer PR-23-AC-62-HSS-EP sensor and Pha

Seite 4

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201312Appendix to the Instructions Manual PR-232.4 Indicating Transmit

Seite 5 - Ultracentrifugation

The Pharma Refractometer PR-23-AC2© Copyright K-Patents 2013. All rights reserved.13Front view Back view186 mm(7.32”)281 mm(11.06”)153 mm(6.04”)304 mm

Seite 6 - 1.1 Design Qualication

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201314Appendix to the Instructions Manual PR-23wireless network (WLAN)

Seite 7 - PR-23-AC

© Copyright K-Patents 2013. All rights reserved.153Installation of PR-23 Pharma Refractometer3.1 Hardware and Software RequirementsPR-23 software is

Seite 8 - 2.2 System components

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201316Appendix to the Instructions Manual PR-23Attaching table top stan

Seite 9

Installation of PR-23 Pharma Refractometer3© Copyright K-Patents 2013. All rights reserved.172Figure 3.2 Sensor stand assembly: Installing the sensor

Seite 10 - 2.3.1 Sensor Model code

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201318Appendix to the Instructions Manual PR-234Figure 3.4 Sensor stan

Seite 11

Installation of PR-23 Pharma Refractometer3© Copyright K-Patents 2013. All rights reserved.193.4 Indicating Transmitter installation for use in table

Seite 12

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 20132Appendix to the Instructions Manual PR-23Do not underestimate or n

Seite 13 - Front view Back view

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201320Appendix to the Instructions Manual PR-233.5 Wiring transmitter

Seite 14

Installation of PR-23 Pharma Refractometer3© Copyright K-Patents 2013. All rights reserved.213.6 Refractometer Instrument VericationThe operational

Seite 15 - Refractometer

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201322Appendix to the Instructions Manual PR-23

Seite 16

© Copyright K-Patents 2013. All rights reserved.2344.1 Ethernet connectionIn addition to software operation via the hardware, the DTR can be consider

Seite 17

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201324Appendix to the Instructions Manual PR-23Figure 4.1 Computer Com

Seite 18

Electronic Data Capture and Storage4© Copyright K-Patents 2013. All rights reserved.25displayed at all times. The IP address of the connected refracto

Seite 19

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201326Appendix to the Instructions Manual PR-23CALC is the calculated c

Seite 20

Electronic Data Capture and Storage4© Copyright K-Patents 2013. All rights reserved.27Figure 4.4 Computer Communication Software CC-11: Data logging

Seite 21

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201328Appendix to the Instructions Manual PR-23The log le is not visib

Seite 22

© Copyright K-Patents 2013. All rights reserved.295Complying with Documentation and Validation Regulations5.1 DocumentationWhen a pharmaceutical comp

Seite 23 - 4.1 Ethernet connection

© Copyright K-Patents 2013. All rights reserved.3ContentsSection 1 K-Patents Products for Vaccine Production and Sucrose Gradient Ultracentrifugati

Seite 24

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201330Appendix to the Instructions Manual PR-235.4 Electronic Data Man

Seite 25 - 4.2.3 Trend View Menu

Complying with Documentation and Validation Regulations5© Copyright K-Patents 2013. All rights reserved.315.9 K-Patents Refractometer System Adherenc

Seite 26 - 4.2.4 Data Logging Menu

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201332Appendix to the Instructions Manual PR-23

Seite 27

© Copyright K-Patents 2013. All rights reserved.336This Installation Qualication (IQ) involves documented verication of the complete system: K-Paten

Seite 28

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201334Appendix to the Instructions Manual PR-236.1.2 Authorized ofcia

Seite 29 - Complying with Documentation

Onsite Qualication Protocols and Records: Installation Qualication6© Copyright K-Patents 2013. All rights reserved.356.2.2 Manufacturers and suppli

Seite 30 - 5.8 System Validation

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201336Appendix to the Instructions Manual PR-236.3.2 DamageDescription

Seite 31

Onsite Qualication Protocols and Records: Installation Qualication6© Copyright K-Patents 2013. All rights reserved.376.5 Operating environmentDescr

Seite 32

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201338Appendix to the Instructions Manual PR-236.7 Setting up the syst

Seite 33

Onsite Qualication Protocols and Records: Installation Qualication6© Copyright K-Patents 2013. All rights reserved.396.8 Electrical connections and

Seite 34 - 6.2 System

Document/Revision No. IM-GB-PR23-ACP: Rev. 1.01 Effective: November 1, 20134Appendix to the Instructions Manual PR-23Section 6 OnsiteQualicatio

Seite 35 - 6.3 IQ Protocol

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201340Appendix to the Instructions Manual PR-236.9 Ethernet connection

Seite 36 - 6.4 Documentation

Onsite Qualication Protocols and Records: Installation Qualication6© Copyright K-Patents 2013. All rights reserved.41Requirement valuesThe correspon

Seite 37 - 6.6 Installation

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201342Appendix to the Instructions Manual PR-236.12 Installation Quali

Seite 38

© Copyright K-Patents 2013. All rights reserved.437Onsite Qualication Protocols and Records: Operational QualicationOperational Qualication (OQ) is

Seite 39

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201344Appendix to the Instructions Manual PR-237.3 Test procedureThe O

Seite 40 - 6.9 Ethernet connection

Onsite Qualication Protocols and Records: Operational Qualication7© Copyright K-Patents 2013. All rights reserved.457.5 System qualicationCheck th

Seite 41

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201346Appendix to the Instructions Manual PR-237.7 Instrument vericat

Seite 42

Onsite Qualication Protocols and Records: Operational Qualication7© Copyright K-Patents 2013. All rights reserved.477.8 Operational Qualication Su

Seite 43

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201348Appendix to the Instructions Manual PR-23Comments (including disc

Seite 44 - 7.4 Authorized ofciator

© Copyright K-Patents 2013. All rights reserved.498After the instrument is qualied, it can be used to measure analytical data. A Standard Operating P

Seite 45 - 7.5 System qualication

© Copyright K-Patents 2013. All rights reserved.51This Instruction Manual Appendix covers K-Patents Pharma Refractometer PR-23-AC when used in the pro

Seite 46 - Refractive Index Liquids

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201350Appendix to the Instructions Manual PR-23

Seite 47

© Copyright K-Patents 2013. All rights reserved.519The need for K-Patents Pharma Refractometer regular maintenance is minimal, due to the construction

Seite 48

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 201352Appendix to the Instructions Manual PR-23

Seite 49 - Routine Operation Phase

© Copyright K-Patents 2013. All rights reserved.5310Other DocumentationYou may want to include the following documents in your les concerning this K-

Seite 50

NotesAppendix to the Instructions Manual PR-23Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.0 Effective: March 1, 201354

Seite 52

www.kpatents.comK-Patents OyP.O. Box 77FI-01511 Vantaa, FinlandTel: +358 207 291 570Fax: +358 207 291 577Email: [email protected], Inc.1804 C

Seite 53 - Other Documentation

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 20136Appendix to the Instructions Manual PR-231.1 Design QualicationD

Seite 54

© Copyright K-Patents 2013. All rights reserved.72The Pharma Refractometer PR-23-AC2.1 System descriptionThe recommended system for the vaccines prod

Seite 55

Document/Revision No. IM-GB-PR23-VACCINES: Rev. 1.01 Effective: November 1, 20138Appendix to the Instructions Manual PR-232.2 System components2.2

Seite 56

The Pharma Refractometer PR-23-AC2© Copyright K-Patents 2013. All rights reserved.91132111589107b7a533a142a71264Figure 2.1 System hardware and softwa